DOLD IP 5880/711 – Insulation monitor
- Brand
- DOLD
- Part number
- IP 5880/711
- Order number
- 0057875
Insulation monitor
Download datasheet (PDF)Ask an engineer
Price and lead time on request — add this part to your quote list and we will reply with pricing and availability.
For insulation monitoring of the IT system of rooms used for medical purposes according to VDE 0100-710.
Features
- According to IEC/EN 61 557-8
- For rooms used for medical purposesaccording to IEC 60364-7-710, DIN VDE 0100-710
- For three-phase and A.C. power systems with 0 … 500 V and 10 … 1000 Hz (IT power systems)
- Adjustable alarm value for ground fault RAL of 50 … 500 kΩ
- Measuring circuit with broken wire protection
- As option, programmable for storing or non-storing of errors
- With reset and test button
- Additional external reset and test buttons can be connected
- LED indicators for operation, insulation fault, and interruption of Measuring circuit
- 2 changeover contacts
- With LED chain for indication of the current insulation status
- IP 5880/711 for connection of the test and display panel UP 5862
- 52.5 mm width
Technical specifications
| Auxiliary voltage | AC |
|---|---|
| Classification | For medically used rooms |
| Earth fault indicator | Yes |
| Enclosure design | Distribution board |
| IMD type | AC |
| Nominal voltage IT system | AC 0 – 500 V |
| Response value | 50 – 500 kΩ |
| Response value type | Adjustable |
| Width | 70 mm |
| Type | IP 5880/711 |
Available variants
| Alarm value | 50 – 500 kOhm |
|---|---|
| Auxiliary voltage | AC 220 – 240 V 50 – 400 Hz |
| Contact equipping | 2 changeover contacts |
| Approvals and Markings | UKCA |
| Variant | For rooms used for medical purposes |